EasyManuals Logo

Hach sc1000 User Manual

Hach sc1000
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
76
Standard Operations
A set of functions is available to set the error and warning status of output modules. Each
of these functions requires a minimum of 2 (or 3) parameters and allows a maximum of
32 parameters. In calculations all functions take the value of first argument A as the
function result, so the use of these functions does not affect the calculated value.
Table 26 Check functions to set errors and warnings
Range Error RNG(A, Min, Max)
When A<Min or A>Max:
Error <E4> “RANGE FUNCTION” is set to the executing card
Range Warning rng(A, Min, Max)
When A<Min or A>Max:
Warning <W1> “RANGE FUNCTION” is set to the executing card
Conditional Error CHK(A, X)
When X is true:
Error <E3> “LOGIC FUNCTION” is set to the executing card
Conditional Warning chk(A, X)
When X is true:
Warning <W0> “LOGIC FUNCTION” is set to the executing card

Table of Contents

Other manuals for Hach sc1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sc1000 and is the answer not in the manual?

Hach sc1000 Specifications

General IconGeneral
Operating Temperature-20 to 60 °C (-4 to 140 °F)
InputsUp to 2 digital or analog sensors
Outputs2 analog, 2 relay
CommunicationModbus
Power Supply100-240 VAC
Enclosure RatingIP66, NEMA 4X