SECTION 5, continued
45
Figure 17 Plumbing Diagram
Table 20 Replacement Tubing Lengths for
Figure 17
Item Description Length (Qty) From... To... Catalog Number
1 1/8" ID, 1/4" OD 4.5 inches (1) Pump Body Out Colorimeter 43293-00
2 1/32" ID, 3/32" OD 7.5 inches (2) Pump Body Out Y-Fitting 44253-00
3 0.042" ID x 0.172 OD 2.0 inches (2) Pump Body In Pump Body Out 54121-00
4 1/16" ID, 3/32" OD 2.0 inches (2) Pump Body In Pump Body Out 42717-00
5 1/32" ID, 3/32" OD 7.0 inches (2) Reagent Bottle Cap Pump Body In 44253-00
6 1/32" ID, 3/32" OD 6.0 inches (2) Reagent Bottle Cap Reagent Vent Fitting 44253-00
7 1/32" ID, 3/32" OD 7.0 inches (2) Reagent Bottle Bottom Reagent Bottle Cap 44253-00
8 1/8" OD, 1/32" wall 48.0 inches (1) Sample Bypass Tee Pump Body In 5189-37
9 1/4" OD x 0.04 W, Black varies (1) Sample Conditioning Out Case Fitting 30616-00
10 1/2" ID varies (1) Instrument Drain Customer Drain (not supplied)
11 1/32" ID, 3/32" OD 1.0 inch (1) Y-Fitting Colorimeter 44253-00
12 0.500" ID x 0.687" OD 12 inches (1) Colorimeter Drain Fitting 54108-00