HAMILTON-C6 Installation Guide
Interaction Panel Mounting Options
Hamilton Medical | HAMILTON-C6 Interaction Panel Mounting Options
19
Required tools
Torx T15 screwdriver for VESA mount assembly
B
Components
Interaction panel mounting clamp options:
1 x Handle mount clamp (PN 160782)
*
* A HAMILTON-C6 Trolley Solution (PN 160029) is required for the installation.
Not included.
** A HAMILTON-C6 ventilator (PN 160021) is required for the installation. Not included.
Note: The interaction panel provides the interface for the VESA mount wall bracket assembly.
VESA mount wall bracket is not included.
A 3-meter (PN 160785) or a 10-meter (PN 160786) VU-IP cable is required for the
pole mount, the rail mount, and the VESA mount. The VU-IP cable is not included.
Clamp A1 must be ordered separately.
Clamp A2 is included in the Pole Mount solution (PN 160031).
Clamp A3 is included in the Rail Mount soluttion (PN 160032).
Safety information
A
A1
Turn off the VU and disconnect the power cable
from the mains power supply before you start the
installation. This will prevent possible electric shock
or damage to the IP or the VU.
Disconnect the VU-IP cable from the rear of the IP
and the rear of the VU before you start the
installation if connected.
A2 1 x Pole mount clamp (PN 160989)
**
A3 1 x Rail mount clamp (PN 160890)
**
1 x VESA adapter plate (PN 10071828)
A1
A2
A3
B
Tests and checks after installation
For details on the electrical safety tests and preoperational
checks on the HAMILTON-C6 ventilator, see Chapter 1 –
Ventilator Unit and Interaction Panel Overview.
Fill in the HAMILTON-C6 ventilator Testing and Acceptance
Report after the initial installation. See Chapter 20.