HAMILTON-C6 Installation Guide
Paramagnetic O2 sensor Assembly
Hamilton Medical | HAMILTON-C6 Paramagnetic O2 sensor Assembly
35
Required tools
Slotted screwdriver (size: # 3 or # 4)
Torx T6 screwdriver
Components
1 x Paramagnetic O2 sensor (PN 160169)
1 x O-ring
1 x Paramagnetic O2 sensor mounting adapter
4 x Countersunk head Torx screws, M2 x 5 mm
(PN PN 420658)
*
4 x Lens head Torx screws, M2 x 5 mm (PN 420393)
**
4 x Delta PT
®
button head Torx screws, M2.2 x 8 mm
(PN 420902 )
***
* Screws for mounting adapter to paramagnetic O2 sensor.
** Screws for mounting the paramagnetic O2 sensor to mounting plate with metal threaded
screw holes.
*** Screws for mounting the paramagnetic O2 sensor to mounting plate with threaded
screw holes in the plastic.
Note: A HAMILTON-C-6 ventilator (PN 160021) is required for the installation. Not included.
Safety information
A
B
C
D
E
F
Turn off the VU and disconnect the power cable from
the mains power supply before you start the
installation. This will prevent possible electric shock or
damage to the VU or the O2 sensor.
Identify the thread type of the mounting plate, then use
the specified screws for the applicable thread, to mount
the paramagnetic O2 sensor to the mounting plate.
If you use the incorrect screws to mount the
paramagnetic O2 sensor, this can cause damage to the
mounting plate.
A
D E
F
B
C
The cable for the paramagnetic O2 sensor connector is
preinstalled and stowed above the sensor compartment.
•
•
Stow the connector cable for the galvanic O2 sensor
above the sensor compartment after the removal of the
galvanic O2 sensor.
Tests and checks afte
r installation
For details on the electrical safety tests and preoperational
checks on the HAMILTON-C6 ventilator, see Chapter 1 –
Ventilator Unit and Interaction Panel Overview.
Fill in the HAMILTON-C6 ventilator Testing and Acceptance
Report after the initial installation. See Chapter 20.