627075/03
160470, 160471
HAMILTON-C6 Installation Guide
DISS/NIST Connector Assembly
Hamilton Medical | HAMILTON-C6 DISS/NIST Connector Assembly
43
Required tools
Torx T10 screwdriver
Components
High-pressure oxygen connector:
1 x DISS connector (PN 160470)
or
1 x NIST connector (PN 160471)
*
2 x Torx screws, M3 x 10 mm
* NIST connector (PN 160471) image shown in overview below.
Note: A HAMILTON-C-6 ventilator (PN 160021) is required for the installation. Not included.
A
B
Safety information
Tests and checks after installation
Turn off the VU and disconnect the power cable
from the mains power supply before you start the
installation. This will prevent possible electric shock
or damage to the VU.
After the installation, connect a high-pressure oxygen hose
to the DISS or NIST connector to check for leakage.
The installation for the DISS connector (PN 160470) and
the NIST connector (PN 160471) are the same.
A
B