HAMILTON-C6 Installation Guide
HAMILTON-H900 Integration Assembly
Hamilton Medical | HAMILTON-C6 – HAMILTON-H900 Integration Kit Assembly
39
Required tools
Allen key (size: 4 mm)
Spanner/wrench: BS 8 mm or 5/16” AF
Components
1 x HAMILTON-H900 integration kit (PN 160783)
Contains:
1 x Power cable
1 x Humidifier communication cabl
(RS232 DB9 serial adapter / 5-pin DIN male plug)
1 x HAMILTON-H900 mounting kit (PN 160969)
Contains:
1 x Hex socket cap screw, M8 x 12 mm
1 x Trolley mounting bracket
1 x M6 slot-nut
1 x X Knurled head thumb screw, M6 x 16 mm
1 x M6 countersunk washer
Mounting kit for third-party humidifier:
1 x Mounting bracket
2 x Hex socket countersunk screws, M5 x 12 mm
Note: A HAMILTON-C6 Trolley Configuration (PN 160029) is required for the installation.
Not included.
HAMILTON-H900 is required for the installation. Not included.
Must be ordered separately. There are different voltage ouputs for the
HAMILTON-H900.
A
A1
A2
B
B1
B2
B3
B4
B5
B6
B7
Safety information
Turn off the VU and disconnect the power cable from
the mains power supply before you start the
installation. This will prevent possible electric shock or
damage to the VU.
A1
A2
B2
A
B
B3
B4
B5
B7
B6
B1
T
e
sts and checks after installation
For details on the electrical safety tests and preoperational
checks on the HAMILTON-C6 ventilator, see Chapter 1 –
Ventilator Unit and Interaction Panel Overview.
Fill in the HAMILTON-C6 ventilator Testing and Acceptance
Report after the initial installation. See Chapter 20.