1
Components
Interaction Panel (IP) / monitor
*
Ventilator Unit (VU)
*
* Must be ordered separately with PN 160021.
Note: The ventilator power cable is country-specific. Order the power cable specific to the
mains power outlet of your country.
A 3-meter (PN 160785) or a 10-meter (PN 160786) VU-IP cable is also available.
A
B
Tests and checks after installation
You must do the electrical safety tests after the initial
installation of the HAMILTON-C6 ventilator.
Configure the HAMILTON-C6 ventilator at the place of
installation according to the institution's or hospital's
protocol.
For additional details on the electrical safety tests and
preoperational checks, refer to the HAMILTON-C6 Service
Manual (PN 627038) and the HAMILTON-C6 Operator’s
Manual (PN 624945).
Fill in the HAMILTON-C6 ventilator Testing and Acceptance
Report after the initial installation. See Chapter 20.
Safety information
Turn off the VU and disconnect the power cable
from the mains power supply before you start the
installation. This will prevent possible electric shock
or damage to the IP or the VU.
B