HAMILTON-C6 Installation Guide
Communication Board Assembly
Hamilton Medical | HAMILTON-C6 Communication Board Assembly
27
Required tools
Slotted screwdriver (size: # 1)
Torx T10 screwdriver
Components
Communication board:
A1 1 x CO2, SpO2, and Aerogen
®
(PN 160184)
Or
A2 1 x CO2 and SpO2 (PN 160185)
2 x Torx screws, M3 x 8 mm
Note: A HAMILTON-C-6 ventilator (PN 160021) is required for the installation. Not included.
Safety information
A
B
Use an applicable electrostatic discharge (ESD) device
such as ESD wrist strap to equalize your body’s charge
with the ventilator unit’s charge before you start the
installation. This will help protect the electronic
components of the ventilator unit and the
communication board from ESD damage.
Turn off the VU and disconnect the power cable
from the mains power supply before you start the
installation. This will prevent possible electric shock
or damage to the VU or the communication board.
A1
A2
A
B
Tests and checks after installation
For details on the electrical safety tests and preoperational
checks on the HAMILTON-C6 ventilator, see Chapter 1 –
Ventilator Unit and Interaction Panel Overview.
Fill in the HAMILTON-C6 ventilator Testing and Acceptance
Report after the initial installation. See Chapter 20.