HAMILTON-C6 Installation Guide
IntelliCuff Kit Assembly
Hamilton Medical | HAMILTON-C6 IntelliCuff Assembly
Required tools
Slotted screwdriver (size: # 3 or # 4)
Components
1 x IntelliCuff pressure controller (PN 951001)
1 x Cuff tubing
1 x USB communication cable
1 x IntelliCuff retaining plate (PN 160834)
Note: A HAMILTON-C6 ventilator (PN 160021) is required for the installation. Not included.
Safety information
A
B
C
D
Do not put batteries in the IntelliCuff battery compart-
ment as the USB communication cable provides
electrical power to the IntelliCuff through the VU.
B
Tests and checks after installation
For details on the electrical safety tests and preoperational
checks on the HAMILTON-C6 ventilator, see Chapter 1 –
Ventilator Unit and Interaction Panel Overview.
Fill in the HAMILTON-C6 ventilator Testing and Acceptance
Report after the initial installation. See Chapter 20.