English |
627075/03
160785, 160786
HAMILTON-C6 Installation Guide
Cablewrap Assembly
Hamilton Medical | HAMILTON-C6 Cablewrap Assembly
23
Turn off the VU and disconnect the power cable
from the mains power supply before you start the
installation. This will prevent possible electric shock
or damage to the IP or the VU.
Required tools
Slotted screwdriver (size: # 2)
Torx T15 screwdriver
Components
1 x Ventilator Unit-Interaction Panel (VU-IP) cable
Available lengths: 3 m (PN 160785) or 10 m (PN 160786)
1 x Cablewrap (PN 160936)
2 x Blanking covers
2 x Torx screws, M4 x 6 mm
Note: A HAMILTON-C-6 ventilator (PN 160021) is required for the installation. Not included.
Safety information
A
B
C
D
D
A
B
C
Tests and checks after installation
For details on the electrical safety tests and preoperational
checks on the HAMILTON-C6 ventilator, see Chapter 1 –
Ventilator Unit and Interaction Panel Overview.
Fill in the HAMILTON-C6 ventilator Testing and Acceptance
Report after the initial installation. See Chapter 20.