HAMILTON-C6 Installation Guide
Testing and Acceptance Report
Hamilton Medical | HAMILTON-C6 Installation Guide – Testing and Acceptance Report
160021
HAMILTON-C6 Ventilator – Testing and Acceptance Report for Initial Installation
SN
__ __ __ __ __ __
SN
__ __ __ __ __ __
____________________________________________
____________________________________________
__ __ __ __ / __ __ / __ __
Task/Check Verify or observe…
HAMILTON-C6 ventilator
Manual Ref.
Task
OK Not OK
Check that the IP-VU cable is connected
correctly to the IP and VU.
Check that the DISS or NIST connector is
installed and a high oxygen supply hose
is connected.
The gas hose connection is
leak-tight.
See the Installation Guide
(PN 627075), Chapter 12; and
the Operator’s Manual
(PN 624945), Chapter 3
Preparing the ventilator.
Check that the ventilator battery is installed.
See the Installation Guide,
Chapter 10.
Connect the breathing circuits.
The breathing circuits are
assembled correctly.
See the Operator’s Manual
(PN 624945), Chapter 3
Preparing the ventilator.
Connect the VU power cable to an AC
power socket.
See the Operator’s Manual
(PN 624945), Chapter 3
Preparing the ventilator.
Do a functional check on the VU
and the IP power cycle (on and off):
Note: Wait for the VU to complete the
start up or shut down process for each
power cycle.
•
Turn on, then turn off the VU with
the push-button switch
: push and
release the switch on the rear of the VU.
•
Turn on, then turn off the IP with the
Power/Standby key
: press and release
the Power/Standby key on the lower
front of the IP.
During the startup process,
the IP alarm lamp flashes
red, yellow, and blue
followed by a short sound
for the speaker test and a
beep for the buzzer test.
See the Operator’s Manual
(PN 624945), Chapter 3
Preparing the ventilator.
See the Installation Guide
(PN 627075), Chapter 4
Shelf Mount Solution Assembly
HAMILTON-C6 Ventilator – Testing and Acceptance Report for Initial Installation [1/2]
Date (YYYY/MM/DD):
Name of installer:
____________________________________________
Name of distributor and country:
Software version:
Interaction panel (IP) serial number:
Ventilator unit (VU) serial number:
Name of hospital and place:
____________________________________________
Do the following tests and checks after the initial installation of the HAMILTON-C6 ventilator:
Do not put the HAMILTON-C6 ventilator into service if any of the tests/checks specified in the Testing and
Acceptance Report fails. Service the ventilator before you put it into service.
59