Task/Check Verify or observe…
OK Not OK
Select the desired language for the user
interface.
Verify that information on
the display is shown in the
selected language.
See the Operator’s Manual
(PN 624945), Chapter 14
Configuration.
Do the tightness test, then calibrate the
flow sensor.
The tightness test passes:
•
•
all breathing circuits’
connections are leak-tight.
The flow calibration is
successful.
See the Operator’s Manual
(PN 624945), Chapter 5
Specifying ventilation settings.
Calibrate the oxygen sensor.
See the Operator’s Manual
(PN 624945), Chapter 5
Specifying ventilation settings.
Do a functional check on the
communication board:
CO2 and SpO2 sensors, and the Aerogen
®
cables
See the Operator’s Manual
(PN 624945), Chapter 14
Configuration and the
Communication board User
Guide
.
Simulate the loss of external power (AC)
alarm test.
The loss of external power
alarm is generated and that
the ventilator is powered by
its backup battery.
See the Operator’s Manual
(PN 624945), Chapter 5
Specifying ventilation settings.
Reconnect the VU power cable to the AC
power socket.
The alarm resets and the
ventilator is powered again
by the AC power.
See the Operator’s Manual
(PN 624945), Chapter 5
Specifying ventilation settings.
Check that the IntelliCuff is setup correctly
if installed.
See the Installation Guide
(PN 627075), Chapter 8; and the
Operator’s Manual
(PN 624945), Chapter 4 Setting
up external devices and sensors.
Check that the HAMILTON-H900 humidifier
is setup correctly if installed.
See the Installation Guide
(PN 627075), Chapter 9 ; and the
Operator’s Manual
(PN 624945), Chapter 4 Setting
up external devices and sensors.
Do the electrical safety tests as specified in
the HAMILTON-C6 Service Manual.
See the Service Manual
(PN 627038), Chapter 6
Electrical safety.
Test Equipment
Electrical safety analyzer
Type/Model Serial Number (SN)
__ __ __ __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __ __ __
Flow analyzer
__ __ __ __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __ __ __
Pressure gauge
__ __ __ __ __ __ __ __ __ __ SN __ __ __ __ __ __ __ __ __ __
See the Operator’s Manual
(PN
624945), Chapter
10
V
entilation settings and
functions.
Verify that the electrical safety
tests are performed
successfully.
Record of test equipment
HAMILTON-C6 Ventilator – Testing and Acceptance Report for Initial Installation [2/2]
HAMILTON-C6 ventilator
Manual Ref.
Task/Check Status: