EasyManuals Logo

Hanna Instruments HI98190 User Manual

Hanna Instruments HI98190
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
26
ISE CALIBRATION HI98191
Press CAL. The primary LCD will displays the ion concentration in the selected unit or “‑‑‑” if not
calibrated and first standard value.
If necessary, press the / keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and within range of the selected standard, the CFM functional key is
displayed.
Press CFM to confirm calibration.
The calibrated value and the second expected standard value will be displayed.
After the first calibration point is confirmed, immerse the ISE electrode approximately
4 cm (1½) into the second calibration solution.
If necessary, press the / keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and within range of the selected standard, the CFM functional key is
displayed.
Press CFM to confirm calibration.
The calibrated value and the third expected standard value will be displayed.
After the second calibration point is confirmed, immerse the ISE electrode approximately
4 cm (1½) into the third calibration solution.
If necessary, press the / keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and within range of the selected standard, the CFM functional key is
displayed.
Press CFM to confirm calibration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98190 and is the answer not in the manual?

Hanna Instruments HI98190 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98190
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals