EasyManuals Logo

Hanna Instruments HI98190 User Manual

Hanna Instruments HI98190
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
47
mV AND TEMPERATURE CALIBRATION (for technical personnel only)
Press CFM to confirm.
The second expected calibrated point is displayed.
Immerse the temperature probe into the second vessel as close as possible to the reference
thermometer. Allow a few seconds for the probe to stabilize.
Use the / keys to set the calibration point value to that of the hot water.
When the reading is stable and within range of the selected calibration point, CFM functional key
is displayed.
Press CFM to confirm. The instrument returns to measurement mode.
Note: Use / keys to change calibration point if necessary (±10.0 °C) around the
point. If the reading is not within range of the selected calibration point, “Wrong
message will blink. Change the temperature probe and restart calibration.
mV CALIBRATION
A two point calibration can be performed at 0 mV and 1800 mV.
Attach to the BNC connector a mV simulator with an accuracy of ±0.1 mV.
Enter the calibration screen. Press mV functional key.
Set 0.0 mV on the simulator.
When the reading is stable and within range of the selected calibration point, the CFM
functional key is displayed.
Press CFM to confirm. The second calibration point of 1800 mV will be displayed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98190 and is the answer not in the manual?

Hanna Instruments HI98190 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98190
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals