mV AND TEMPERATURE CALIBRATION (for technical personnel only)
Set 1800.0 mV on the simulator.
When the reading is stable and within range of the selected calibration point, the CFM functional
key is displayed.
Press CFM to confirm. The instrument returns to calibration screen.
Press ESC to return to measurement mode.
Notes: If the reading is not within range of the selected calibration point, “WRONG” tag will
blink. Verify calibration condition or contact your vendor if you cannot calibrate.
Press CAL or ESC in any moment of the calibration process. The instrument will return in
the measurement mode.