SETUP
The following table lists the specific range parameters.
Item Description Valid value Default
Calibration Timeout
(pH & ISE)
Number of days after
Calibration warning is
displayed
Disable, 1 to 7 days Disable
First point mode (pH) Management of
1 point calibration
Replace or offset Replace
Custom buffer (pH) Custom buffer setting Max. 5 buffers No
View calibration
points (pH)
Display calibration
points
Enable or disabled Enable
Display out of Cal.
Range Warning
Enable or disabled Enable
ISE probe
(HI98191 only)
Type of ISE probe Custom or Standard (17) Fluoride
ISE unit
(HI98191 only)
User, ppt, g/L, ppm, mg/L,
ppb, µg/L, mg/mL, M,
mol/L, mmol/L, % W/V
ppm