27
CALIBRATION
RESTORING FACTORY CALIBRATION
Restoring factory calibration will delete the user calibration and restore the factory default calibration.
Use the / keys to highlight Restore factory calib. and then press Select.
8.5. ATMOSPHERIC PRESSURE CALIBRATION
The HI98199 is factory calibrated for atmosphere pressure. The user can perform a single point
calibration or restore the factory calibration.
From the Calibration menu, use the / keys to highlight Atm. Pressure and then press Select.
• Select the Custom pressure option.
• Use the alphanumeric keypad to enter the known value from the reference meter and then press
Accept to confirm.
• The stability counter will count down and the message “Ready” and Confirm will be displayed.
Press Confirm to store the calibration point.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press Measure to return to the measurement screen.
• Press OK to return to the Calibration menu.
RESTORING FACTORY CALIBRATION
Restoring factory calibration will delete the user calibration and restore the factory default calibration.
Use the / keys to highlight Restore factory calib. and then press Select.