CALIBRATION
RESTORING FACTORY CALIBRATION
Restoring factory calibration will delete the user calibration and restore the factory default calibration.
Use the / keys to highlight Restore factory calib. and then press Select.
8.5. ATMOSPHERIC PRESSURE CALIBRATION
The HI98199 is factory calibrated for atmosphere pressure. The user can perform a single point
calibration or restore the factory calibration.
From the Calibration menu, use the / keys to highlight Atm. Pressure and then press Select.
• Select the Custom pressure option.
Use the alphanumeric keypad to enter the known value from the reference meter and then press
Accept to confirm.
The stability counter will count down and the message “Ready” and Confirm will be displayed.
Press Confirm to store the calibration point.
After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press Measure to return to the measurement screen.
• Press OK to return to the Calibration menu.
RESTORING FACTORY CALIBRATION
Restoring factory calibration will delete the user calibration and restore the factory default calibration.
Use the / keys to highlight Restore factory calib. and then press Select.