41
TROUBLESHOOTING / ERROR MESSAGES
13. TROUBLESHOOTING / ERROR MESSAGES
HI98199 displays error messages to aid in troubleshooting. Warnings are displayed for noncritical
issues, while Errors are displayed for critical issues. See the calibration chapter for messages that can
occur during calibration.
Other messages are listed below.
• “Log space full”: appears when the meter memory is full and additional data cannot be logged.
Delete one or more lots from the meter.
• “Power fault. Check the probe cable”: appears when powering up the meter with a probe connected
and the meter detects a high load on the probe connection. Check the probe cable. If the problem
persists, contact your local Hanna Instruments Office.
• “Language data not available”: appears when powering up the meter, if the language file is not
seen by the meter. Restart the meter. If the problem persists, contact your local Hanna Instruments
Office.
• ‘’Dead meter battery!’’: appears if the batteries are too low to power the meter and will
automatically turn off. Replace the alkaline batteries to continue.