(Fig. 814-4)
After entering parameter setting interface of Time mode, press to choose
parameter item to be edited, it will be marked with blue rectangle background. Press OK
key, the blue rectangle disappears, then the current editable position will be marked
white rectangle background, input the value according to (Fig. 814-3), VTBI and Time
are adjustable, after confirming, press OK to save.
Rate will be automatically calculated by system according to VTBI and Time, and
shows on the screen.
Note: If the rate is larger than limits, after pressing Start, it will remind " Flow rate
is out of range, please reset the parameters". Here please press to return the
parameter setting interface to reset the parameters.
Return: If want to reselect the infusion mode, press to return main menu
( Fig.814-1 ).
(3)Weight Mode