33
pump fails.
Battery running out
alarm (In cases
when only battery
supplying can be
used at power
failure or in moving
process)
1. At least 3min before the
pump turns itself off
automatically for the battery
running out.
Stop using and charge the battery full.
2. The battery is aging or the
charging circuit of syringe
pump fails.
Please ask the distributor or
manufacturer to repair.
Occlusion alarm
1. the syringe is blocked in
operation process
Press Clear key to eliminate alarm.
2. The blockage sensitivity is
too high.
Refer to 8.4.4 of this Manual and
adjust the blockage pressure alarm
value of syringe pump.
3. The sensor of syringe pump
fails
Please contact the distributor or
manufacturer to repair.
Near completion
alarm
The cumulative injection
volume approaches the preset
amount.
Press Clear to eliminate alarm sound
and wait until injection finishes. This
alarm can be set off, refer to 8.4.10 of
this Manual for details.
Injection
completion alarm
Preset injection volume is
completed
Press Clear to eliminate alarm sound,
and press Stop to eliminate alarm.
Near empty alarm
Medicine inside syringe is
about to be pushed empty
Press Clear to eliminate alarm sound
and wait until the liquid is pushed
empty.
Empty alarm
Medicine inside syringe has
been pushed empty
Press Clear to eliminate alarm.
Syringe off alarm
The push handle of syringe has
not been pushed onto the
Press Clear to eliminate alarm. Please
install the syringe again correctly.