34
syringe, and the syringe pump
fails to recognize syringe
Handle off alarm
Clutch is pressed during the
operation process of syringe
pump.
Press Clear to eliminate alarm, and
stop pressing the clutch and start to
inject again.
Syringe hand-push
off
The hand-push of syringe does
not fit well with the push-pull
box of syringe pump.
Press Clear to eliminate alarm. Please
install the syringe again correctly.
Abnormal syringe
roll hem
The roll hem of syringe is not
installed inside the slot
correctly.
Press Clear to eliminate alarm. Please
install the syringe again correctly.
E0, E1
E2, E3
1.E0: data communication error
Restart the syringe pump, confirm
parameters and restart injection. If the
alarm is given again, contact with the
manufacturer to inspect and repair.
2. E1: transmission part failure
Restart the syringe pump, confirm
parameters and restart injection. If the
alarm is given again, contact with the
manufacturer to inspect and repair.
3.E2: The Syringe Pump’s
motor has problem
Reboot the machine and load the
parameters of last infusion to try
operation again. If problem still occurs,
contact distributor /manufacturer for
repair
4.E3: The Syringe Pump’s data
storage has problem
Reboot the machine and load the
parameters of last infusion to try
operation again. If problem still occurs,
contact distributor /manufacturer for
repair