35
Abnormal
AC fail alarm
AC power plug
off and battery exhausted durin
g infusion, the pump will
alarm in sound and red light for
3 minutes
please charge the battery fully.
Note:The set parameters (including alarm setting) are stored in memory chips. The saved data will
not be lost after power failure and can be saved for more than 10 years. Alarm setting will be called out
when the machine is turned on again.
Method to verify the normal operation of alarm system:
Note: The verification below cannot be carried out when the infusion pipe is connected to patients;
otherwise, patients will be risky.
1). AC power-down alarm
Unplug AC power cord after turning the pump on, the device will display AC fails and give alarm
sound.
2). Operation forgotten alarm
The device will display “No Operation” and give alarm sound if no operation is made after 2min when
the infusion pump is in non-operation and non-alarm status.
3). Low battery alarm
The device will display low battery and give alarm sound when battery voltage is lower than 7.1V.
4). Battery exhausted alarm
The device will display battery exhausted, stop operation and give alarm sound when battery voltage
is lower than 6.65V.
5). Occlusion alarm
When infusion set is blocked artificially in operation process, the pressure indicator will increase to
the right and the device will display occlusion and give alarm sound after a period of time (depends on flow
rate).
6). “Almost done” alarm
If setting as ON, “Almost Done” alarm shall sound 3 minutes before VTBI is complete.
7). Finish alarm
“Finish” alarm shall sound when VTBI is complete