syringe pump should not be put close to or stacked on other devices. If it has to be put close to or
stacked on other device, it should be observed and verified that if it can operate normally under its
application configuration;
Class A device is planned to be used in industrial environment. Due to the conducted disturbance
and radiated disturbance of syringe pump, it may be difficult to ensure EMC in other environment;
The application of accessories and cables, other than the cables sold by syringe pump manufacturer
as the spare parts of its internal components & parts, can increase pump emission or reduce
interference immunity.
Attachments:
Guidance and Manufacturer’s Declaration - EME
The syringe pump intended for use in the electromagnetic environment specified below. The
customer of the user of the syringe pump should assure that it is used in such an environment
Electromagnetic environment – guidance
Radio frequency
emission
GB 4824
The syringe pump use RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Radio frequency
emission
GB 4824
The syringe pump is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage