Guidance and Manufacturer’s Declaration-Electromagnetic Immunity
The syringe pump is intended for use in the electromagnetic environment specified below. The
customer or the user of syringe pump should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conformance Level
Electromagnetic
Environment
Guidance
Electrostatic
discharge
IEC 61000-4-2
±8 kV contact
discharge
±15 kV air discharge
±8 kV contact discharge
±15 kV air discharge
The floor should be
covered with wood,
concrete or tiles; if it is
covered by composite
materials, then the
relative humidity should
be at least 30%.
Electrical fast
transient burst
IEC 61000-4-4
±2kV topower line
±1kV toinput/output line
±2kVtopower line
Network power should
have the quality that is
typical for application in
commercial environment
or hospital.
power supply network that supplies buildings used for
domestic purposes
Harmonic emission
GB 17625.1
Not applicable
Voltage
fluctuation/Flash
radiation
GB 17625.2
Not applicable