55
No operation alarm
Low priority
Yellow,steady
Table 2 Alarm conditions and alarm signal delay
Alarm classification
Alarm Condition Delay
Alarm Signal Delay
Handle off alarm
10ms
100ms
Syringe off alarm
10ms
100ms
Roll hem abnormality alarm
10ms
100ms
Syringe push off
10ms
100ms
Occlusion alarm
840s@1ml/h
27s@25ml/h
100ms
Low battery alarm
10ms
100ms
Battery exhausted alarm
500ms
100ms
Empty
10ms
100ms
Almost empty
10ms
100ms
Almost Done alarm
10ms
200ms
Finished alarm
10ms
200ms
AC fail alarm
10ms
200ms
No operation alarm
120ms
200ms