56
Table 3 Characteristic parameters of alarm signals
Table 4 Occlusion response characteristic
The above test uses “Boon” brand of 5ml syringe. All the data are obtained by using Boon” brand
syringe.The syringe pump has pressure release function. When occlusion alarm sounds, the pressure in
the infusion line system will release automatically, so the bolus volume could be neglected when occlusion
block release.
Flow rate
(ml/h)
Occlusion
alarm level
Occlusion
pressure(KPa)
Occlusion alarm
time
1
Low
41
12min59sec47s
Middle
104
35min29sec
High
144
42min53sec
5
Low
54
4min37sec69s
Middle
90
7min09sec
High
153
9min50sec