T210
LOAD-CARRYING CAPACITY
16
The rotational speed reference signal Z is additionally outputted on pin H. It is also a 5 V
square wave signal. At 1 pulse per revolution, it is suitable for measuring high rotational
speeds, as this places lower demands on the evaluation unit.
Information
The T210 torque transducers are suitable for a nominal (rated) rotational speed of max.
30,000 rpm depending on the nominal (rated) measuring range.
Tip
To ensure an optimum distribution of grease in the bearing, turning the measurement shaft
once before the start of the measurement with the maximum expected rotational speed is
recommended.