Protocol description ASCII, SECS/HSMS
Customer manual
Version 1.2
LFM LP Reader 15.01.2019
Page 60 of 119
HERMOS AG
Gartenstrasse 19 Tel. +49(9279) 991-0
D-95490 Mistelgau Fax. +49(9279) 991-100
1)
: The use of the sensor message B depends on internal parameter 49.
Further internal parameters (from parameter 16 onwards) are available.
Please ask the manufacturer HERMOS if you need anything.
If an acknowledgement is not sent by the host, the device
repeats the message according to the set value. Only then is
an error message sent
00 - endless
01 .. 99 – Number of attempts
Default: 3
Activates / deactivates the sensor event messages A or B to
the host, that the input sensor has been opened or closed.
Bit0 of the Watchport parameter (event message A):
0 - deactivate message A
1 - activate message A when sensor opened/released
Bit1
1)
of the Watchport parameter (event message B):
0 - deactivate message B
1 - activate message B when sensor closed/occupied
Default: 01
Address of the first antenna port (0 ... E)
Default: 1