17/34
The lid can not be opened
Error
The lid can not be opened
The rotation display indicates "Lid closed" and the lid is closed.
Error cause
Solenoid on lid lock is defective
Loosen contact on plug S404 / Pin 1 and 3
Power board (A1) defective
Jumper for lid lock control is set on power board (A1), see
chapter 5.5
Measurement Remove Plug S404 and measure between Pin 1 and 3:
Solenoid ok: 5.7 k
Solenoid defective:
See also section 9.2.4.
Error code reset Release the lid by using the emergency release.
No display
Error No power supply on control board, no display
Error
consequence
No operation possible
Error cause
No mains supply
Power board (A1) defective
Control panel (A3) defective
Flat ribbon cable to control panel (A3) defective
Overtemperature fuse F3 defective
Measurement 1. Mains supply power board A1 S402L to S402N
2. Power board A1 S501 Pin 1 (GND) to S501 Pin 4 (+5V)
3. Control panel A3 S502 Pin 1 (GND) to S502 Pin 4 (+5V)
4. Overtemperature fuse F3. Remove plug S405 and measure
directly at both flat connections of the overtemperature fuse:
Fuse OK : 0
Fuse defective: . See also section 9.2.3.