4.4 Defects without Error indications
No speed indication / Machine-Version-Error
Error The centrifuge stops after few seconds and the speed display
extinguished and in the time display appears the set machine
version.
Error cause
Wrong machine version adjusted
Example for indication:
RPM x 100 t
Maschinenversion / Machine version
START
STOP
Proceed as the following: 1. Adjust the correct machine version by
using the and arrow keys in the
time area.
ROTOFIX 32 = 4
MIKRO 20 = 2
EBA 20 = 1
2. Press the key
STOP
in order to store
the set machine version.
3. Perform a MAINS RESET.
The lid can not be opened
Error The lid can not be opened.
The rotation display indicates "Lid opened", but the lid is closed.
Error cause
Micro switch at lid lock is defective
Loosen contact on plug S404 / Pin 5 and 6
Power board A1 defective
Jumper for lid lock control is set on power board (A1), see
chapter 5.5
Measurement Remove plug S404 and measure between Pin 5 and 6:
Switch closed: 0
Switch opened:
See also section 9.2.4.
Error code reset
Release the lid by using the emergency release.