8.6 Overtemperature fuse F3
Remove the two cables from the overtemperature fuse.
Unscrew the fastening screws.
Exchange the overtemperature fuse.
8.7 Flat ribbon cable
Unplug the flat ribbon cable from the power board.
Disassemble the control board from the housing(see section 8.4).
Remove the three fastening screw on back side of the control panel.
Lift off carefully the panel of the housing.
Unplug the flat ribbon cable from the control board.
Assemble in reverse order.
8.8 Appliance plug with mains switch
Press the upper and the lower fastening clip and remove the appliance plug out of
the housing
Remove the plugs and the earth connection from the appliance plug.
Remove the fuse holder, see chapter 8.9.
Exchange the appliance plug.
8.9 Exchange mains fuse
The fuse holder is placed in the appliance plug.
Press the clip upward.
Extract the fuse holder.
Exchange the defective fuse.
8.10 Imbalance switch S2
Unplug S503 from the power board.
Remove the to fastening screws.
Remove the cables at the imbalance switch.
Exchange the complete imbalance switch.
After mounting the imbalance switch, adjust it as described in chapter "Imbalance
switch-off".
8.11 Lid lock Y1
Remove the two fastening screws on the top of the upper housing.
Unplug the cable from the switch and the solenoid.
Remove the earth connection from the lid lock.
Exchange the lid lock.