25/34
8.6 Overtemperature fuse F3
• Remove the two cables from the overtemperature fuse.
• Unscrew the fastening screws.
• Exchange the overtemperature fuse.
8.7 Flat ribbon cable
• Unplug the flat ribbon cable from the power board.
• Disassemble the control board from the housing(see section 8.4).
• Remove the three fastening screw on back side of the control panel.
• Lift off carefully the panel of the housing.
• Unplug the flat ribbon cable from the control board.
• Assemble in reverse order.
8.8 Appliance plug with mains switch
• Press the upper and the lower fastening clip and remove the appliance plug out of
the housing
• Remove the plugs and the earth connection from the appliance plug.
• Remove the fuse holder, see chapter 8.9.
• Exchange the appliance plug.
8.9 Exchange mains fuse
• The fuse holder is placed in the appliance plug.
• Press the clip upward.
• Extract the fuse holder.
• Exchange the defective fuse.
8.10 Imbalance switch S2
• Unplug S503 from the power board.
• Remove the to fastening screws.
• Remove the cables at the imbalance switch.
• Exchange the complete imbalance switch.
• After mounting the imbalance switch, adjust it as described in chapter "Imbalance
switch-off".
8.11 Lid lock Y1
• Remove the two fastening screws on the top of the upper housing.
• Unplug the cable from the switch and the solenoid.
• Remove the earth connection from the lid lock.
• Exchange the lid lock.