EasyManua.ls Logo

Heyl Testomat 2000 - Operating and Display Elements (Continued); Understanding Display Functions and Indicators

Heyl Testomat 2000
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Functions of the operating and
18
Display functions
1 Limit value status displays (red/green)
The display 1 illuminates red if limit value 1 has been reached or
exceeded. The display 1 illuminates green if the value falls below
the limit value. The same principle applies to limit value 2 and
display 2. Behaviour is completely opposite with switch function 4.
With switch function 5 both displays illuminate green if the
measured value is within the range of LV1 and LV2.
2 Text display (4 lines)
Displays the current analysis result as well as all important
statuses and programming data.
2a = the current measured value for measuring point 1 (M1) and
2 (M2) is displayed in line 2 and 3, e.g. M1: 17.0 mgl/l
Value exceeds the measuring range = ">" e.g. M1:
>50.0 mgl/l
2b = the set limit values LV1 and LV2 are displayed in line 4
3 Alarm (red)
Indicates malfunctioning/error message or warning message.
4 Maintenance message (yellow)
Indicates current maintenance requests
5 Status display of the active instrument components (line)
Six displays indicate the current instrument and analysis status
Cancelling error messages/warning messages
Press to acknowledge the message and, if necessary,
eliminate the cause of the fault.
All error and
warning messages
are alternately
shown in line 1 of
the standard
display!
Inlet
valve
Analysis
running
Outlet
valve
Dosing
message
Without
function
Analysis
stop
2a 2b
Polymer
M1: 20.2 mg/l
1:0025.6 2:0015.4 mg/l
NOTE

Table of Contents