2.2 The Head Up Function Does Not Operate
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed, Page 2 - 7
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
8. Replace the controls on the siderail with the faulty function (refer to
procedure 4.21 on page 4-46). The head up function operates from this
siderail.
Yes No
↓→Replace the siderail with the faulty function (refer to procedure
4.16 on page 4-37). If this corrects the problem, go to “Final
Actions” on page 2-4. Otherwise, call Hill-Rom Technical
Support at 800-445-3720 (USA).
9. Replace the control box (refer to procedure 4.1 on page 4-3). If this
corrects the problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 800-445-3720 (USA).