Chapter 7 Troubleshooting
―――――――――――――――――――――――
If no power is supplied:
If you're using a battery, check that it has sufficient
remaining power. (See 2.1: Preparations )
If you're using the AC adapter, check that it's fully
inserted into the AC adapter terminal and socket.
If no power is supplied by the AC adapter but the battery
works, the adapter may be defective. (Make sure that
you're using the 9445-02 or 9445-03 AC ADAPTER.)
The counter doesn't become zero:
If this occurs in DC A, AC+DC A or DC V modes, use
the auto-zero adjustment function (2.5.1).
If this occurs in AC A, AC V or AC+DC V mode, use the
zero-cancel correction function (2.5.2).
If this occures in AC + DC A mode after performing the
suto-zero-adjustment, perform the zero cancel correction
function (2.5.2).