Attentions During Use
―――――――――――――――――――――――
WARNING
.
Do not input voltage exceeding 600 Vrms (1000 V
max).
To avoid electric shock when replacing the
batteries, first disconnect the clamp portion from
the object to be measured. Also, before using the
product after replacing the batteries, replace the
cover and screw.
When replacing the batteries, be sure to insert
them with the correct polarity. Otherwise, poor
performance or damage from battery leakage
could result.
To avoid the possibility of explosion, do not short
circuit, disassemble or incinerate batteries.
Handle and dispose of batteries in accordance
with local regulations.
To prevent electric shock, when measuring the
voltage of a power line use a test lead that
satisfies the following criteria:
Conforms to safety standards IEC61010 or
EN61010
Of measurement category III or IV
Its rated voltage is higher than the voltage to be
measured
The test leads provided with this instrument
conform to the safety standard EN61010.
Use a test lead in accordance with its defi ned
measurement category and rated voltage.
Do not use the instrument where it may be exposed to
corrosive or combustible gases. The instrument may
be damaged or cause an explosion.