Chapter 2 Measurement Procedure
――――――――――――――――――――――――
NOTE
NOTE
2.5.2 Zero-Cancel Correction Function
2.6 Data Hold Function HOLD
If you press
0ADJ/RESET
again during the auto-
zero-adjustment in the internal circuit, the auto-zero-
adjustment is canceled.
Do not perform auto-zero adjustment when using
auto range in the DC V mode. Always switch to the
range you will use (a manual range) before
performing the adjustment.
Use the zero-cancel correction function when the
counter fails to become zero under no input in AC
A, AC+DC A, AC V, or AC+DC V mode.
1. Press
HOLD
to display
HOLD
annunciator.
2. Press
0ADJ/RESET
.
ADJ
annunciator blinks.
Do not perform a zero-cancel correction when using
the auto range. Always switch to the range you will
use (a manual range) before performing the
adjustment.
When there is an input or the counter decreases, the
measured values will be evaluated lower by pressing
0AJD/RESET
.
If the counter is zero, the zero-cancel correction
function does not work.
In AC+DC A mode, the zero-cancel correction
function does not work, unless auto-zero-adjustment
is complete.
The zero-cancel correction function works only on
the counter. It will not calibrate output values.
This function freezes the counter at any desired
point for easy reading.