35
―――――――――――――――――――――――――――
Chapter 2 Measurement Procedure
――――――――――――――――――――――――
NOTE
NOTE
2.5.2 Zero-Cancel Correction Function
2.6 Data Hold Function HOLD
・ If you press
0ADJ/RESET
again during the auto-
zero-adjustment in the internal circuit, the auto-zero-
adjustment is canceled.
・Do not perform auto-zero adjustment when using
auto range in the DC V mode. Always switch to the
range you will use (a manual range) before
performing the adjustment.
Use the zero-cancel correction function when the
counter fails to become zero under no input in AC
A, AC+DC A, AC V, or AC+DC V mode.
1. Press
HOLD
to display
HOLD
annunciator.
2. Press
0ADJ/RESET
.
ADJ
annunciator blinks.
・Do not perform a zero-cancel correction when using
the auto range. Always switch to the range you will
use (a manual range) before performing the
adjustment.
・ When there is an input or the counter decreases, the
measured values will be evaluated lower by pressing
0AJD/RESET
.
・ If the counter is zero, the zero-cancel correction
function does not work.
・ In AC+DC A mode, the zero-cancel correction
function does not work, unless auto-zero-adjustment
is complete.
・ The zero-cancel correction function works only on
the counter. It will not calibrate output values.
This function freezes the counter at any desired
point for easy reading.
1. Press
HOLD
.
HOLD
annunciator lights on the