Chapter 7 Troubleshooting
――――――――――――――――――――――――
Symptom Battery Battery clip Test leads
Unit does not come
on.
Yes Yes
indication appears
and unit immediately
turns off.
Yes
indication
appears.
Yes
Unit turns off during
use.*
Yes Yes
Voltage cannot be
measured.
Yes
Remedy:
If problem persists,
request service.
Replace battery. Check
connection of
battery to clip.
Check test leads
for broken wire.
NOTE
Chapter 7
Troubleshooting
If the unit seems not to be working normally, check
the following points first before requesting service.
* When APS (auto power-off) is effective, the unit is
automatically shut down when no key is pressed for
about 10 minutes. (See 2.9, "Auto Power-Off
Function.")
Repair is required if E.001 to E.005 is indicated on
the counter, when power is turned on.