Chapter 2 Measurement Procedure
――――――――――――――――――――――――
NOTE
MAX AVEMIN Instantaneous value
(no annunciator)
2.2.5 Out
p
ut Function
In case that the counter doesn't become zero under
no input in peak measurement mode, even though
you pressed
0ADJ/RESET
to reset the peak data, the
clamp sensor may be magnetized. Quit the peak
measurement mode, and perform the auto-zero
adjustment by
0ADJ/RESET
. Then make the
settings again. (A few counts would remain,even if
you push
0ADJ/RESET
.)
The hold value does not change, unless a larger
value is measured, but be careful to avoid accidental
loss of data resulting from the auto power-off
function. (See 2.9: Auto power-off function.)
Use the REC function to make measurements longer
than the auto power OFF time.
You cannot output peak values. When pressing
OUTPUT
during peak measurement, the present
measured value is outputted.
To check transitional peak value, press
MAX/MIN
to shift to the Instantaneous value (no annunciator).
An output of AC/DC 1 V is produced for 2000-
count on the full scale of each current range.
Select either
REC
(record output) or
MON
(waveform output).
(In DC A mode, only
MON
is available. See 1.5:
Modes)
1. Press
RANGE
to set the current range.
2. Press
OUTPUT
.
REC
or
MON
annunciator lights
and activates the output function, automatically
disabling the auto power-off function. (
APS
annunciator is turned off.)