Chapter 2 Measurement Procedure
――――――――――――――――――――――――
NOTE
2.9 Auto Power-Off Function APS
・ When you need minimum value and average value
data, make sure to activate the recording function
during measurement. If the function is activated
when there is no input, the minimum value will
remain zero. Also, when deactivating the recording
function, press the
HOLD
key to terminate
measurement once the minimum value and average
value data have been read. If you disconnect the
clamp or test lead from the circuit under
measurement without deactivating the recording
function beforehand, the minimum value will be
zero.
・ When the unit is turned off, accumulated data are
lost.
・ When the
APS
annunciator is displayed, the auto
power-off function is active.
・ If no key is pressed for about 10 minutes, the unit
turns itself off automatically.
・ Immediately before turning off automatically,
APS
annunciator blinks and a beep tone is heard for
about 30 seconds.
・ By pressing any key except
POWER
,youwill
extend the powered state for another 10 minutes.
Procedure for disabling the auto power-off
function.
・ Press
POWER
with holding down
HOLD
,when
you turn power on.
・ Use the recording function (REC) by pressing
MAX/MIN
.
・ Press
OUTPUT
in a current mode.