Chapter 7 Troubleshooting
――――――――――――――――――――――――
○ If no power is supplied:
・ If you're using a battery, check that it has sufficient
remaining power. (See 2.1: Preparations )
・ If you're using the AC adapter, check that it's fully
inserted into the AC adapter terminal and socket.
・ If no power is supplied by the AC adapter but the battery
works, the adapter may be defective. (Make sure that
you're using the 9445-02 or 9445-03 AC ADAPTER.)
○ The counter doesn't become zero:
・ If this occurs in DC A, AC+DC A or DC V modes, use
the auto-zero adjustment function (2.5.1).
・ If this occurs in AC A, AC V or AC+DC V mode, use the
zero-cancel correction function (2.5.2).
・ If this occures in AC + DC A mode after performing the
suto-zero-adjustment, perform the zero cancel correction
function (2.5.2).