61
―――――――――――――――――――――――――――
Chapter 7 Troubleshooting
――――――――――――――――――――――――
The measured value is smaller than the estimated
value.
Current measurement:
Check that the clamp sensor is firmly closed.
Check that the frequency of the measured circuit is within
the range provided in the specifications. (A smaller value
will be displayed for a high inverter carrier frequency.)
Check that the proper steps have been taken, according to
procedure described in 2.5.1: Auto-zero adjustment
function and 2.5.2: Zero-cancel correction function.
Check that you're using the proper mode. (See 1.5:
Modes.)
Check that the peak value is below the circuit dynamic
value provided in the specifications. (To check
transitional peak values, see Notes in 2.2.4 and 2.3.4:
Peak hold measurement.)
Check that the crest factor (peak value / RMS value) is
below the circuit dynamic value provided in the
specifications.
Check that the battery warning annunciator is off.