47
―――――――――――――――――――――――――――
Chapter 3 Specifications
――――――――――――――――――――――――
Current Specifications
Maximum permissible
current
2000 Arms continuous, 2840 Amax.
See Fig.4
Effect of conductor
position
within ±0.7% (in any direction from
sensor center)
External magnetic
field interference
AC 400 A/m (external magnetic fields)
corresponds to 2 A or less (display)
Maximum rated
voltage to earth
max. 600 Vrms
2500
2000
1500
1000
500
0
10 100 1k 10k 100k
Frequency [Hz}
Fig.4 Frequency-dependent deletion characteristics