63
―――――――――――――――――――――――――――
Chapter 7 Troubleshooting
――――――――――――――――――――――――
The measured value is larger than the estimated
value.
Current measurement:
Check that you're using the proper range.
Examine the waveform using
MON
output function to
confirm that no components but the estimated frequency
are being used.
Check that the counter has been reset with
0ADJ/RESET
before peak measurement.
Look for magnetic fields, electrical fields or possible noise
sources near the unit.
The measured value would be bigger than the actual value
in case that the power supply changes the battery to the
AC adapter such as back from a blackout. (See Chapter 5:
AC Adapter (Optional) )
Voltage measurement:
Check that you're using the proper range.
Check that the counter has been reset with
0ADJ/RESET
before peak measurement.
Look for magnetic fields, electrical fields or possible noise
sources near the unit.
Frequency measurement:
Look for magnetic fields, electrical fields or possible noise
sources near the unit.
Check the waveform. Some special frequencies can't be
measured, such as those of inverters.