53
Measuring Current (DT4252, DT4253, DT4256)
3.9 Measuring Current
(DT4252, DT4253, DT4256)
DC/AC current is measured.
DANGER
• Do not input any voltage to the current measurement
terminals. Doing so may result in short circuit
accidents.
• To avoid electrical accidents, turn off the power to
the circuit before measuring and then connect the
test leads.
Measuring DC/AC Current
Function
• µA Selected to measure 600.0 µA DC or less. (DT4253)
• mA Selected to measure 60.00 mA DC or less. (DT4253)
The % conversion of 4-20 mA can be checked in the sub display.
• A Selected to measure 10 A DC/AC or less. (DT4252, DT4256)
The % conversion of 4-20 mA can be checked in the sub display.
(Only the DT4256)
When measuring an unknown current
Set to the high range (mA for the DT4253).