Retaining the Measured Value
Conditions for auto holding
Function
Stable range for auto
holding
(Display count)
Threshold for auto
holding
(Dead zone display count)
AC voltage 120 or less (except 1000 V
range)
20 or less (1000 V range)
120 or less (except 1000 V
range)
20 or less (1000 V range)
DC voltage
*
2
120 or less (except 1000 V
range
*
1
)
20 or less (1000 V range
*
1
)
120 or less (except 1000 V
range
*
1
)
20 or less (1000 V range
*
1
)
AUTO V 120 or less 120 or less
Continuity check 100 or less 4900 or more
Resistance 100 or less 4900 or more
Diode 40 or less 1460 or more
AC (Clamp) 50/100/25/50/100/25/50 or
less (each range)
50/100/25/50/100/25/50 or
less (each range)
DC (µA) 120 or less 120 or less
DC (mA) 120 or less 120 or less
DC (A) 120 or less (except 10 A
range)
20 or less (10 A range)
120 or less (except 10 A
range)
20 or less (10 A range)
AC (A) 120 or less (except 10 A
range)
20 or less (10 A range)
120 or less (except 10 A
range)
20 or less (10 A range)
*1: DT4254 is 1500 V range.
*2: No function is available for the mV range.