Accuracy Table
5
AUTO V
Range
Accuracy
*1
Input impedance
DC, 40 to 500 Hz Over 500 Hz to 1 kHz
600.0 V ±2.0% rdg. ±3 dgt. ±4.0% rdg. ±3 dgt. 900 k
Ω
±20%
*2
1800 k
Ω
±20%
*3
Overload protection: 1100 V DC (1870 V DC
*3
)/1100 V AC or 2×10
7
V • Hz
(energized for 1 minute)
Transient overvoltage: 8000 V
Crest factor: The crest factor is 3 up to 4000 counts and reduces linearly to
2 at 6000 counts.
Connection method: DC coupling
*1: For AC voltage, the accuracy is specied in 1% or more of the range,
however, ±5 dgt. should be added to 5% or less of the range.
Accuracy guarantee range for frequency: 40 Hz to 1 kHz (Measured
values outside the accuracy guarantee range for frequency are also
displayed.) The accuracy is not specied for strain waveforms outside
the range of 40 Hz to 1 kHz.
For 100 Hz with the lter ON, ±1.5% rdg. is added to the accuracy
specication between 40 Hz and 100 Hz and the accuracy is not specied
in 100 Hz or more.
For 500 Hz with the lter ON, ±0.5% rdg. is added to the accuracy
specication between 40 Hz and 500 Hz and the accuracy is not specied
in 500 Hz or more.
*2: DT4253, DT4255, DT4256
*3: DT4254