3.10 Other Functions
110
The following operations are not available when in the maximum value or minimum value hold state:
Range switching
The instrument will display [Err.16] if any range key is pressed. (p.179)
Similarly, the range will not change during auto-range operation. The range will be fixed to the range
at the time the HOLD key was pressed.
Setting changes (integration time, number of averaging iterations, VT/CT ratios, etc.)
The setting parameter will light up on the Settings screen, and you will not be able to change it.
The flashing cursor cannot be moved to parameters that cannot be changed while in the hold state.
To change a setting, press to return to the instantaneous value display.
For parameters other than waveform peak values, the maximum and minimum values are compared using
the absolute values of measured values. For waveform peak values, the maximum value for the waveform
is shown as the maximum value, and the minimum value of the waveform is shown as the minimum value.
For integration times, integrated values, and time average values, no maximum values or minimum values
are determined. For integration times and integrated values, instantaneous values are directly displayed.
For time averages, [- - - - - ] is displayed.