3.2 Configuring Settings
48
Selecting the desired range
Press the range key to select the desired range. The lamp for the selected range key will light up, and the dis-
play value will change to reflect the selected range.
3.2.4 Selecting the Voltage and Current Ranges
When input exceeds 1,000 V, ±1,500 Vpeak, or 70 A, ±100 Apeak
The maximum input voltage and maximum input current are 1,000 V, ±1,500 Vpeak, and
70 A, ±100 Apeak. If the maximum input voltage or maximum input current is exceeded,
immediately halt measurement, deactivate power to the measurement lines, and dis-
connect the cables from the instrument. Continuing measurement with maximum input
exceeded will damage the instrument and cause bodily injury.
Do not input a voltage or current in excess of each measurement range. Doing so will damage
the instrument.
When using a wiring mode other than 1P2W, the voltage and current ranges will be standardized using the
CH1 settings.
Allow the following amount of time to elapse before reading measured values after changing the range:
When the frequency of input set to the synchronization source is 10 Hz or greater
Approx. 0.6 s
When the frequency of input set to the synchronization source is less than 10 Hz
See: "3.2.7 Setting the Timeout" (p.55)
Channels for which the range has changed will use the invalid data display [- - - - -] until the data is
updated.
When measuring frequencies of 10 Hz or less, it is necessary to set the timeout to a value other than 0.1 sec.
See: "3.2.7 Setting the Timeout" (p.55)
When using external current sensor input as the current input method, only the 10 A, 20 A, and 50 A keys
are valid for selecting the current range. Pressing another range key will cause TYPE1 or TYPE2, indicat-
ing the type of external current sensor, to be displayed, and the range will not be changed.
Ranges cannot be changed while integration is being performed or during display hold or maximum value/
minimum value hold operation.
Default settings: 300 V voltage
20 A current
CH1
Voltage range Current range