3.4 Viewing Harmonic Measured Values
71
The instrument displays the results of harmonic analysis for each channel’s voltage, current, and active
power. Since all calculation processing is performed in parallel internally, you can obtain harmonic measured
values with simultaneity with other measured values simply by switching display parameters.
Additionally, when the synchronization frequency is 45 Hz to 66 Hz, the instrument can perform harmonic
measurement that complies with IEC 61000-4-7:2002.
Set the synchronization source for the harmonic measurement to be performed with the instrument as
described in "3.2.5 Setting the Synchronization Source (SYNC)" (p.51). The source can be set separately for
each wiring mode.
The following table summarizes the instrument’s harmonic measurement parameters and corresponding dis-
play methods:
Switching display modes
3.4 Viewing Harmonic Measured Values
3.4.1 Setting the Synchronization Source
3.4.2 Method for Displaying Harmonic
Measurement Parameters
Display mode
Normal display
LEVEL lamp off
HD% lamp off
Harmonic level display
LEVEL lamp on
HD% lamp off
Harmonic content percentage
display
LEVEL lamp off
HD% lamp on
Measurement
items
Total harmonic voltage distortion
Total harmonic current distortion
Fundamental wave voltage RMS value
Harmonic current RMS value
Fundamental wave active power
Fundamental wave apparent power
Fundamental wave reactive power
Fundamental wave power factor
Fundamental wave voltage/current phase
difference
Inter-channel voltage fundamental
wave phase difference
Inter-channel current fundamental
wave phase difference
Harmonic voltage RMS value
Harmonic current RMS value
Harmonic active power
0th to 50th order
Harmonic voltage content
percentage
Harmonic current content
percentage
Harmonic active power
content percentage
0th to 50th order
1 Press .
2 You can check the display mode
based on the LEVEL lamp and HD%
lamp state.
Example: LEVEL lamp on
LEVEL lamp HD% lamp
Normal display Off Off
Harmonic level display On Off
Harmonic content percentage display Off On