5.3 Measurement specifications
145
Voltage Waveform Peak Value Measurement Specifications
(Upk: Displayed as V pk on panel display)
Constraints During integration operation and while integration is stopped, the following operational
constraints apply until integrated values are reset:
: Setting and changes allowed -: Setting and changes not allowed
Item
During integration operation or while integration is
stopped
Measurement range
-
(Fixed to range when integration was started.)
Current input method switching -
Wiring -
Frequency measurement range
(Zero-Crossing Filter)
-
Timeout -
Integration time -
Synchronization source -
Zero Adjustment
-
(Can be adjusted while integration is stopped.)
D/A output parameters
Number of averaging iterations -
VT ratio -
CT ratio -
RS-232C transfer speed
GP-IB address
LAN
Harmonic analysis order upper limit value -
Measurement method Measures the voltage waveform’s peak value (for both positive and negative polarity)
based on sampled instantaneous voltage values.
Range configuration
Measurement accuracy ±2.0% f.s. at DC and when 10 Hz f 1 kHz (f.s.: voltage peak range). Provided as
reference value when 0.1 Hz f < 10 Hz and when in excess of 1 kHz.
Effective measuring range ±5% to ±100% of voltage peak range (up to ±1,500 V)
Display range ±0.3% to ±102% of voltage peak range (values less than ±0.3% are subject to zero-sup-
pression)
Over-range display Displays [o.r] when input exceeds ±102% of the voltage peak range.
Invalid data display Upksum is displayed as [- - - - -].
Voltage range 15 V 30 V 60 V 150 V 300 V 600 V 1000 V
Voltage peak range 90.000 V 180.00 V 360.00 V 900.00 V 1.8000 kV 3.6000 kV 6.0000 kV